8.6 Using the Analog Trigger Function
────────────────────────────────────────────────────
.6.4 Period Trigger (RMS Recorder Function Excluded)
NOTE
NOTE
The period trigger setting determines the period reference voltage and period
range. This trigger is tripped when the measured period of the rise (or fall) of
the reference voltage exceeds the period range.
・The setting for the period range of the period trigger changes depending on
the sampling period.
Lower limit: Ten times the sampling period or more
Upper limit: 20000 times the sampling period or less
・In the recorder function, check the sampling period setting when using the
period trigger.
Method Screen: STATUS 2
1. Select the period trigger.
(1) Call up the STATUS 2 screen.
(2) Move the flashing cursor to the position
shown in the figure on the left.
(3) Use the function keys to select cycle.
2. Set the period range.
(1) Move the flashing cursor to the position
shown in the figure on the left.
(2) Use the
JOG control or the function keys to
set the lower limit of the period.
: Value up, large step
: Value up, small step
: Value down, small step
: Value down, large step
(3) Set the upper limit in the same way.
On the DISPLAY screen, the setting cannot be
made with the
JOG control.