182
────────────────────────────────────────────────────
8.6 Using the Analog Trigger Function
────────────────────────────────────────────────────
8
.6.2 Window-In, Window-Out Trigger (RMS Recorder Function
Excluded)
Upper lim
it
Lower lim
it
上限値
下限値
Window-in-trigge
r
Window-out-trigge
r
Upper lim
it
Lower lim
it
Window-in trigger
Set upper limit level and lower limit level and activated when the input signal
enters the range between these limits.
Window-out trigger
Set upper limit level and lower limit level and activated when the input signal
leaves this range.
Method Screen: STATUS 2, DISPLAY
1. Select the window-in or window-out trigger.
(1) Call up the STATUS 2 or DISPLAY screen.
(2) Move the flashing cursor to the position
shown in the figure on the left.
(3) Use the function keys to select window-in or
window-out.