.2 Setting the Trigger Mode
・The trigger mode determines the way triggering is used to control operation of
the 8835-01.
・When all trigger sources are OFF, a recording operation begins immediately
(free-run operation).
Method Screen: STATUS 2, DISPLAY
1. Call up the STATUS 2 or DISPLAY screen.
2. Move the flashing cursor to the trig mode
item, as shown in the figure on the left.
3. Use the function keys to make the selection.
: SINGLE
Trigger is registered only once. After
START key was pressed, unit starts
waveform recording when triggering occurs
and continues for preset recording length.
Measurement then ends automatically.
: REPEAT
Trigger is registered continuously. Unit is
in trigger standby condition when trigger
conditions are not met. Measurement ends
when STOP key is pressed.
: AUTO
Trigger is registered continuously. If trigger
conditions are not met within 1 second,
waveform recording starts automatically
and continues for preset recording length.
Measurement ends when STOP key is
pressed (memory recorder function only).