Maximum electrical charge input is 500 pC at range 6 high sensitivity
and 50000 pC at range 6 low sensitivity. Make sure input is within the
range to avoid damage to unit.
When selecting Preamp mode and starting up, confirmation message
appears. Check channel connection before proceeding with measuring.
When measuring starts in Preamp mode, electric current (2 A, + 15 V)
output is active. Switching to another mode (Electrical Charge or
Voltage) terminates output. (Not terminated by
STOP key.)
NOTE
Setting the Measurement Mode
Method Screen: CHANNEL 1
1. Call up the CHANNEL 1 screen.
2. Use the
JOG control or the function keys to
select Charge or Preamp.
: Move the cursor up in the selection window.
: Move the cursor down in the selection
window.
Adjusting Sensor Sensitivity
・Adjust sensor sensitivity.
・Sensor sensitivity is measured as a specific value of the acceleration sensor.
・Sensor sensitivity modifies measurement range. Adjust sensor sensitivity
before setting measurement range (see Section 15.2.7).
Method Screen: CHANNEL 1
1. Move the flashing cursor to the point shown
in the figure on the left.
2. Use the function keys or the jog control to
make the setting.
: Value up, large step
: Value up, small step
: Value down, small step
: Value down, large step
Sensor sensitivity is measured as a specific value of the acceleration sensor.
Adjust sensor sensitivity to sensor to assure accurate acceleration values.