341
────────────────────────────────────────────────────
15.5 Measurement Errors Caused by Signal Source Internal Resistance
────────────────────────────────────────────────────
Es:Signal voltage
Rs: Signal source resistanc
e
Rin: Input resistance
                 Rin
Measurement errorsEs 1− ――――― V
                RsRin
R
s
Ri
n
E
s
1
5.5 Measurement Errors Caused by Signal Source
Internal Resistance
If the signal source impedance is higher than the input impedance of the unit,
a measurement error will occur.
The input impedance of the 8936 ANALOG UNIT is 1 MΩ. If the signal
source impedance is 1 kΩ, an error of about 0.1% will occur.