11.5 Vernier Function
────────────────────────────────────────────────────
Displays the magnified waveform.
×
Displays the waveform with no
vernier scaling.
Displays the compressed waveform
.
NOTE
The actual captured waveform Adjusted to 2 Vp-p using the vernier function
.
1
1.5 Vernier Function
Using fine adjustment, the input voltage can be matched to a desired reading.
For example, an actual input voltage of 1.8 V can be converted to a 2.0 V
reading.
Method Screen: DISPLAY
1. Press the DISP key to call up the DISPLAY
screen.
2. Move the flashing cursor to the position
shown in the figure on the left.
3. Use the function keys to select the vernier
function.
Use
to magnify and to compress. The
adjustment range is from 1/2 to 2 times of
the original waveform.
: Cancels the vernier function. (The original
waveform returns.)
: Magnifies the waveform.
: Compresses the waveform.
The vernier function is not applicable to a waveform after waveform
processing*. *: Advanced version
Example: Changing a 1.8 Vp-p waveform to a 2.0 Vp-p waveform