199
────────────────────────────────────────────────────
8.9 Using the External Trigger Function
────────────────────────────────────────────────────
External trigger connector
.9 Using the External Trigger Function
・An external signal can be used as trigger source.
・The EXT TRIG connector is used for this purpose.
・The external trigger is activated by either shorting the EXT TRIG terminal or
applying a falling edge signal going below 2.5 V.
Method Screen: STATUS 2
1. Press the STATUS key to call up the
STATUS 2 screen.
2. Move the flashing cursor to the position
shown in the figure on the left.
3. Use the function keys to make the selection.
: External trigger is not used.
: External trigger is used.