8.8 Using the Timer Trigger Function
────────────────────────────────────────────────────
stop
NOTE
Relation between timer trigger and AND/OR linking
(1) When trigger sources are set to OR
All trigger sources are valid. If other trigger sources have been set, triggering
can also occur before the start time or after the stop time.
(2) When trigger sources are set to AND
・Measurement is carried out from the start time to the end time. Triggering
occurs at the preset intervals if the conditions for the other trigger sources
are also met at these points.
・If the interval has been set to 0s, triggering occurs at any point between the
start and stop time, if if the conditions for the other trigger sources are met.
・ When the trigger mode has been set to SINGLE, only one triggering action
occurs after the measurement was started.
・ To perform recording at regular intervals, establish the following settings.
Trigger mode: continuous
Trigger source linking: AND or OR
Other trigger sources: all OFF