4.5 Start and Stop Measurement Operation
────────────────────────────────────────────────────
LED
Lights during measuremen
t
4
.5 Start and Stop Measurement Operation
Method
1. Press the START key to initiate measurement or set the unit to trigger
standby.
2. Press the
STOP key during measurement to stop the measurement.
(see Section 3.1.4.)
When the
STOP key is pressed once, triggering occurs, waveforms are
displayed and printed out, and measurement is stopped.
When the STOP key is pressed twice, the 8835-01 is forcibly stopped.
When measurement is forcibly stopped following trigger generation,
waveform data collected up to the point of termination remains after
measurement is stopped.
There will be no waveform data if measurement is forcibly stopped prior to
trigger generation. However, during continuous measurement, any
waveform data remaining from a prior trigger will be displayed.
When the auto save function is ON, waveform data collected up to the point
of termination is automatically stored.