7.6 Making the Settings of the 8939 STRAIN UNIT
────────────────────────────────────────────────────
NOTE
NOTE
.6.4 Auto-balancing
Method 1 Screen: CHANNEL 1, DISPLAY
1. Call up the CHANNEL 1 or DISPLAY screen.
2. Move the flashing cursor to the point shown
in the figure on the left.
3. Use the
JOG control or the function keys to
make the selection.
: Move the cursor up in the selection window.
: Move the cursor down in the selection
window.
On the DISPLAY screen, the selection window
is not displayed.
Method 2 Screen: CHANNEL 1, DISPLAY
Press the INPUT RANGE key to set the range for each channel.
・The
INPUT RANGE key can be used regardless of where the flashing cursor
is located, if the CHANNEL 1 or DISPLAY screen is displayed.
・If the variable function is enabled, the size of a waveform on the screen does
not change, even if the measurement range is changed.
・This function aligns the reference output level of the adapter with the
specified origin position.
・Allow the unit to warm up for at least approx. one hour to ensure that the
internal temperature of the input units has stabilized.
・It can be specified whether the auto balancing is applied only to a selected
channel or to all channels.