9.3 Special Function Settings [ SETUP ]
────────────────────────────────────────────────────
9
.3.12 Setting PRINT/EXT.SMPL
NOTE
Selecting external printing/external sampling terminal.
System reset defaults to external printing.
PRINT: Enables external printing in all functions. By setting time axis range
(frequency in FFT) to External in memory recorder and FFT (advanced
version), it functions as external sampling terminal.
EXT.SMPL: Functions as external sampling terminal. (Memory recorder and
FFT only)
Method Screen: SYSTEM 1 (SET UP)
1. Press the SYSTEM key to call up the SET
UP screen.
2. Move the flashing cursor to the position
shown in the figure on the left.
3. Use the function keys to make the selection.
: Use for external printing and external
sampling terminal.
: Use for external sampling terminal.
In memory recorder and FFT (advanced version), setting time axis range
(frequency in FFT) to External determines use of external sampling terminal
even when PRINT is selected.
For external printing, set time axis range (or frequency) to another setting
other than External.