33
────────────────────────────────────────────────────
2.9 Notes on Measurement
────────────────────────────────────────────────────
.9.1 Using a Voltage Transformer
8835-01
GND terminal
H
L
PT
Voltage transformer
(PT) ground termina
Ground
8835-01
GND terminal
PT
Inpu
When the voltage transformer has no ground termina
When the voltage transformer has a ground termina
.9.2 Maximum Input Voltage
8835-0
8938
H
L
370 VAC, DC
400 VDC max
H
L
30 Vrms or 60 VDC
30 Vrms or 60 VDC
30 Vrms o
30 Vrms or 60 VDC
30 Vrms o
60 VDC
When making measurements on an AC power line for example, using a
voltage transformer, be sure to connect the voltage transformer ground
terminal to ground.
8936 ANALOG UNIT, 8938 FFT ANALOG UNIT
8937 VOLTAGE/TEMP UNIT, 8946 4 ch ANALOG UNIT
8939 STRAIN UNIT