34
────────────────────────────────────────────────────
2.9 Notes on Measurement
────────────────────────────────────────────────────
8835-0
*
*: +, - 500 pC (at range 6 high sensitivity),
+, - 50000 pC (at range 6 low sensitivity
DC DC+AC
8940 F/V UNIT
8947 CHARGE UNIT
Difference between "370 V AC, DC" and "400 V DC max." indication
370 V AC, DC: Rms value is displayed.
400 V DC max.: Instantaneous value is displayed.
The maximum input voltage (400 V DC max.) is defined as the superposition
of DC component and AC peak, as shown in the figure below.