.3.2 Start Key Backup
・Assign channels to allocate memory for measurement data.
・The setting is valid only when the memory recorder function is used.
・The smaller the number of channels in use, the longer will be settable
recording length (see Section 14.3).
Method Screen: SYSTEM (SETUP) 11-07
1. Press the SYSTEM key to call up the SET UP
screen.
2. Move the flashing cursor to the position shown
in the figure on the left.
3. Press the function key to make the setting.
: Allocates the memory for measurement data
to channel 1. Measurement cannot be made
on channels 2 to 8. Only logic channel A can
be used.
: Allocates the memory for measurement data
to channels 1 and 2. Measurement cannot be
made on channels 3 to 8. Only logic channels
A and B ca be used.
: Allocates the memory for measurement data
to 4 channels. Measurement cannot be made
on channels 5 to 8. All logic channels can be
used.
: Allocates the memory for measurement data
to 8 channels. All logic channels can be used.
・When this function is enabled, the unit will operate as follows: If the power
supply is interrupted during recording operation (while the
START key LED is
lit), and then the power supply is restored, the 8835-01 goes back into the
measurement operation mode, and recording is restarted immediately.
・If the trigger function is used, the unit goes into trigger standby mode.
However, this is disabled with the timer trigger.
Method Screen: SYSTEM 1 (SET UP) 11-08
1. Press the SYSTEM key to call up the SET
UP screen.
2. Move the flashing cursor to the position
shown in the figure on the left.
3. Press the function key to make the setting.
: The start key backup function is not used.
: The start key backup function is used.