20
────────────────────────────────────────────────────
2.4 Probe Connection
────────────────────────────────────────────────────
1
2
5
6
3
4
7
8
2
.4.2 8937 VOLTAGE/TEMP UNIT
WARNIN
G
Never connect the connection cable to the 8835-01 while the connection
cable is already connected to the measurement object. Otherwise there
is a risk of electric shock.
A common GND is used for voltage and temperature input on all
channels. Never input voltage and temperature simultaneously, since
doing so could result in damage to the sample being tested.
When an uninsulated thermocouple is used to measure temperature at a
point carrying electric potential, take care not to touch the terminals.
Otherwise there is a risk of electric shock.
The voltage and temperature input and the 8835-01 frame are insulated.
CAUTIO
N
When disconnecting the BNC connector, be sure to release the lock before
pulling off the connector. Forcibly pulling the connector without releasing the
lock, or pulling on the cable, can damage the connector.
Use a screwdriver or similar to attach and remove the thermocouple.
Setting Channels in 8946 4 ch ANALOG UNIT
When using 8946 4 ch ANALOG UNIT, input channels are not numbered in
numeric order as 1,2,3,4. Channel number is determined by connection to
units. (See below.) To verify position of channels, see Channel Guide.
Connection cable connection (Voltage measurement)
For measuring voltage with 8937, use 9198 CONNECTION CORD.
1. Align the BNC connector with the guide groove of the unit input connector,
and turn clockwise while pressing in to lock the connector.
2. To remove from the unit, turn the BNC connector counterclockwise to release
the lock, then pull it.