NOTE
Thermocouple connection (Temperature measurement)
1. Strip off the insulation as shown in the illustration.
2. Push the tab with a screwdriver or similar.
3. While keeping the tab depressed, insert a stripped thermocouple into the
connector openings.
4. Release the tab to lock the thermocouple.
5. While keeping the tab depressed, remove the thermocouple.
・The press-button type terminal board of the 8937 VOLTAGE/TEMP UNIT is
only for connection to a thermocouple. Do not use thermocouples other than
the specified types (K, J, E, T, N, R, S, B).
・If the thermocouple is connected in reverse, the temperature reading will
not be correct.
・If the temperature input terminal is exposed to a strong draft, loss of
thermal equilibrium at the input may result in measurement error. When
taking measurements under such conditions, arrange the unit in such a
manner that the input terminal is protected for direct exposure to drafts.
・If ambient temperature changes suddenly, loss of thermal equilibrium can
result in measurement error. When this occurs, allow the unit to acclimate
to the new temperature for about one hour, then take measurements after
thermal equilibrium is reached.