7.6 Making the Settings of the 8939 STRAIN UNIT
────────────────────────────────────────────────────
NOTE
7
.6.5 Setting the Magnification/Compression Ratio Along the
Measurement Range
Method Screen: CHANNEL 1, DISPLAY
1. Call up the CHANNEL 1 or DISPLAY screen.
2. Move the flashing cursor to the point shown
in the figure on the left.
3. Use the function keys to make the selection.
: Carries out auto-balancing for the selected
channel only.
: Carries out auto-balancing for all of the
strain unit channels.
Connect the sensor, make sure there is no input, and execute auto
balancing.
Auto-balancing cannot be performed during measurement.
During auto-balancing, the keys are inactive.
Repeat the auto-balancing when the setting range was changed, when the
input unit was changed, when the strain gauge adapter was changed, when
the zero position was changed, when the power is on/off, or when the system
is set to reset.
When there is a sudden change in ambient temperature, the zero position
may drift. To assure continued measurement precision, perform the auto-
balancing again.
Specifies the magnification/compression ratio for each channel to be used for
display and recording.
Performs magnification/compression using the center of the screen as
reference.
For the setting, see Section 7.3.5.