172
────────────────────────────────────────────────────
8.3 Setting the Pre-trigger (Memory Recorder and RMS Recorder Functions)
────────────────────────────────────────────────────
(1) In the Memory recorder function
5%
95%
Example
recording range (recording length)
50%
50%
95%
Trigger point
50%:
50% before and 50% after trigger
95%:
95% before and 5% after trigger
-95%:
R
ecording begins at point passing 95%
of full recording length after trigger.
5%:
5% before and 95% after trigger
(2) In the RMS recorder function
Example
    
recording range (recording length)
20 DIV
5 DIV
10 DIV
Trigger point
0 DIV:
R
ecords full recording length after
trigger.
5 DIV:
R
ecords 5 DIV before trigger and
r
emaining (recording length-5) afte
r
trigger.
10 DIV:
R
ecords 10 DIV before trigger and
r
emaining (recording length-10)
after trigger.
8
.3 Setting the Pre-trigger (Memory Recorder and RMS
Recorder Functions)
The pre-trigger function serves to record the waveform not only after but also
before triggering has occurred.
In the memory recorder function, using the recording start point as 0% and
the recording end point as 100%, the trigger point can be specified in percent.
In the RMS recorder function, using the recording start point as 0 DIV, the
trigger point can be specified in DIV (0, 5, 10 DIV).