140
────────────────────────────────────────────────────
7.6 Making the Settings of the 8939 STRAIN UNIT
────────────────────────────────────────────────────
7
.6.6 Setting the Zero Position
7
.6.7 Setting the Low-Pass Filter
The position of 0 με is set.
For the setting, see Section 7.3.6.
Low-pass filters internal to the input units are set.
Effective for removing unneeded high-frequency components.
Method Screen: CHANNEL 1, DISPLAY
1. Call up the CHANNEL 1 or DISPLAY screen.
2. Move the flashing cursor to the point shown
in the figure on the left.
3. Use the function keys to make the selection.
: No low-pass filter is connected.
: Connect a filter with a cutoff frequency of 10
Hz
: Connect a filter with a cutoff frequency of 30
Hz
: Connect a filter with a cutoff frequency of
300 Hz
: Connect a filter with a cutoff frequency of 3
kHz

Table of Contents