181
────────────────────────────────────────────────────
8.6 Using the Analog Trigger Function
────────────────────────────────────────────────────
In the recorder function
: Trigger filter is disabled.
: Trigger filter is enabled.
Filter width is 10 ms.
Example
To cause triggering at point A or point B with the sine wave shown below, make
the following settings.
Point A trigger level: 200 mV, trigger direction (slope): rising ( )
Point B trigger level: -600 mV, trigger direction (slope): falling ( )