2.1 Installation of the Unit
────────────────────────────────────────────────────
1
2
3
4
5
6
7
8
9
10
11
12
13
14
A
Ventilatio
NOTE
Ambient conditions
Temperature 5to40℃,23±5℃ recommended for high-precision
measurements
Humidity 35 to 80%RH (no condensation); 50±10%RH (no
condensation) recommended for high-precision measurements
Ventilation Take care not to block the ventilation openings and assure
proper ventilation. When using the unit in an upright
position, take care not to block the openings on the bottom, as
it could overheat and be damaged, or cause a fire.
Avoid the following locations:
・Subject to direct sunlight
・Subject to high levels of dust, steam, or corrosive gases
(Avoid using the equipment in an environment containing corrosive gases
(e.g., H
2
S, SO
2
,NI
2
, and CI
2
) or substances that generate harmful gasses
(e.g., organic silicones, cyanides, and formalins)).
・Subject to vibrations
・In the vicinity of equipment generating strong electromagnetic fields