Software Reference 199
V4.2 LabChip GX User Manual PerkinElmer
Excluded Peaks Tab (Continued)
• For Aligning check box - If selected, the data signal is
realigned so the selected excluded peaks match the specified
size or time. Note that incorrect alignment settings can cause
analysis errors.
• Exclude All - If selected, all peaks in the specified range are
excluded from analysis.
• Apply to Wells - Specifies the wells that the excluded peak is
applied to. <All> specifies that the excluded peak applies to all
wells. Clicking on the column opens the Select Wells window to
choose the specific wells to apply the excluded peak to.
Export button - Opens the Export Excluded Peaks window to
create an Excluded Peaks File (.gep) from the current settings in
the tab. Import the .gep file into another assay to use the same
excluded peaks in another assay.
Restore button - Restores the settings to the last saved settings
for excluded peaks.
Import button - Opens the Import Excluded Peaks window to
select an Excluded Peak File (.gep) to import. The .gep file is
created by exporting excluded peaks from another assay.
Clear button - Deletes all Excluded Peaks from the table.