Troubleshooting and Diagnostics 289
V4.2 LabChip GX User Manual PerkinElmer
Current Leakage Check Failed
Warning Message: “Current leakage check failed.”
Chip Temperature Warning
Warning Message: “Average chip temperature is out of range.”
Possible Causes Recommended Actions
Dirty electrode block. Clean the O-rings (see page 281).
Run the Current Leak Diagnostic Test (see page 291).
If the problem is not resolved, contact PerkinElmer
Technical Support (see page 3).
Possible Causes Recommended Actions
Chip heater failed. Restart the instrument. If problem persists, contact
PerkinElmer Technical Support (see page 3).