Troubleshooting and Diagnostics 291
V4.2 LabChip GX User Manual PerkinElmer
Diagnostics
The LabChip GX software contains a set of diagnostic tests to verify
proper operation of the LabChip GX or GX II instrument. These
tests are not dependent on assay chemistry. All tests must pass to
ensure the instrument is functioning properly. To open the System
Diagnostics Window, select Validation Diagnostics.
The following procedures are included in this section:
• Running the Diagnostics Tests
• “Description of Diagnostic Tests” on page 292
Running the Diagnostics Tests
To begin running the Diagnostics tests, click the Run Tests button
on the System Diagnostics Window. The tests run from top to
bottom, prompting you to load the correct chips or plates as
required for each test. Follow the prompts. As each test is
completed, the results display on the right side of the window.
Supplies Required to Run Diagnostics Tests:
• Test Chip A (PerkinElmer P/N 760453)
• Test Chip B (PerkinElmer P/N 760489)
• Test Chip C (PerkinElmer P/N 760624)
• Test Plate with barcode label
To order Test Chip A, B, or C, contact PerkinElmer Technical
Support (see page 3).