LabChip GX Instrument Description 279
V4.2 LabChip GX User Manual PerkinElmer
Assay Voltage
Chip Pressure
Chip Temperature Control
Fluorescence Detection
Light Source (Red laser diode)
Barcode Reader
Minimum Voltage/Current -100V (limited by current)
Maximum Voltage/Current -3000V (limited by current)
Pressure Range +50 to -5 psig
Temperature Range Ambient + 5°C to 40°C (no active cooling)
Accuracy ± 0.5°C
Thermal limit switch 65°C - 70°C
Detection Wavelength Bandpass 670-725 nm
Data rates 20, 30, 40, 50, 60, 80, 100, and 120 Hz
Warmup Time 30 seconds
Wavelength 635 nm
Power output 7.5 mW
Barcode Engine Microscan MS3CCD
Supported Barcode
Types
Code 39, Code 93, Code 128