EasyManua.ls Logo

PerkinElmer LabChip GX - Page 227

PerkinElmer LabChip GX
346 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Software Reference 227
V4.2 LabChip GX User Manual PerkinElmer
Optics Calibration Window (Continued)
New Correction Factor - The correction factor calculated from the
instrument (Reference Contrast / Scan Results Contrast).
Warm and Scan/Scan/Abort button -
Warm and Scan: Warms up the chip and laser, focuses the
chip, and measures the optical contrast.
Scan: Scans without the warmup (available only after a scan is
complete or aborted, while laser and chip are still warmed up).
Abort: stops the scan or warm currently in progress.
Apply button - Saves the New Correction Factor to the
instrument’s non-volatile flash memory.
Done button - Closes the Optics Calibration window.

Table of Contents

Related product manuals