Software Reference 227
V4.2 LabChip GX User Manual PerkinElmer
Optics Calibration Window (Continued)
New Correction Factor - The correction factor calculated from the
instrument (Reference Contrast / Scan Results Contrast).
Warm and Scan/Scan/Abort button -
• Warm and Scan: Warms up the chip and laser, focuses the
chip, and measures the optical contrast.
• Scan: Scans without the warmup (available only after a scan is
complete or aborted, while laser and chip are still warmed up).
• Abort: stops the scan or warm currently in progress.
Apply button - Saves the New Correction Factor to the
instrument’s non-volatile flash memory.
Done button - Closes the Optics Calibration window.