Operation 39
V4.2 LabChip GX User Manual PerkinElmer
Adding a New Plate (Continued)
The instrument will move to the specified number of wells in
each corner of the plate, punching holes in the tape.
The Punch test is acceptable if the punched hole is >0.5mm
from the edge of the well.
Holes will never be perfectly centered.
Possible sources of errors:
Sipper splay
Plate tolerances
Robot alignment error
If the holes are too close to the edge (<0.5mm), adjust the X-
Margin or Y-Margin values in the Add Plate window.
7 Select the new plate name in the Start Run Window when
starting the run.