Remote control commands
R&S
®
SMA100B
471User Manual 1178.3834.02 ─ 09
:CSYNthesis:OTYPe <Mode>
Defines the shape of the generated clock signal.
Parameters:
<Mode> SESine | DSQuare | CMOS | DSINe
SESine = single-ended sine
DSINe = differential sine
DSQuare = differential square
CMOS = CMOS
*RST: SESine
Example: See Chapter 14.7, "CSYNthesis subsystem", on page 469.
Manual operation: See "Output Type" on page 305
:CSYNthesis:FREQuency
<Frequency>
Sets the frequency of the generated clock signal.
Parameters:
<Frequency> float
Numerical value
Sets the frequency
UP|DOWN
Varies the frequency step by step.
The frequency is increased or decreased by the value set with
the command :CSYNthesis:FREQuency:STEP.
Range: 100E3 to 1.5E9
Increment: 0.001
*RST: 10E6
Example: See Chapter 14.7, "CSYNthesis subsystem", on page 469.
Manual operation: See "Frequency" on page 306
:CSYNthesis:POWer <Power>
Sets the power level of the generated clock signal.
Parameters:
<Power> float
Numerical value
Sets the level
UP|DOWN
Varies the level step by step.
The level is increased or decreased by the value set with the
command :CSYNthesis:POWer:STEP[:INCRement].
Range: -24 to 10
Increment: 0.01