Remote control commands
R&S
®
SMA100B
668User Manual 1178.3834.02 ─ 09
[:SOURce<hw>]:LIST:MODE...........................................................................................670
[:SOURce<hw>]:LIST:POWer..........................................................................................671
[:SOURce<hw>]:LIST:POWer:POINts?.............................................................................671
[:SOURce<hw>]:LIST:TRIGger:EXECute..........................................................................672
[:SOURce<hw>]:LIST:TRIGger:SOURce.......................................................................... 672
[:SOURce<hw>]:LIST:RUNNing?..................................................................................... 673
[:SOURce<hw>]:LIST:DWELl <Dwell>
Sets the global dwell time. The instrument generates the signal with the frequency /
power value pairs of each list entry for that particular period.
See also Significant parameters and functions.
Parameters:
<Dwell> float
Range: 1E-3 to 100
Increment: 1E-6
*RST: 0.01
Example: See Example "Create an RF list and activate the list mode"
on page 665.
Manual operation: See "Global Dwell Time" on page 203
[:SOURce<hw>]:LIST:DWELl:MODE <DwelMode>
Selects the dwell time mode.
Parameters:
<DwelMode> LIST | GLOBal
LIST
Uses the dwell time, specified in the data table for each value
pair individually.
GLOBal
Uses a constant dwell time, set with command [:
SOURce<hw>]:LIST:DWELl.
*RST: GLOBal
Example: See Example "Create an RF list and activate the list mode"
on page 665.
Manual operation: See "Dwell Time Mode" on page 203
[:SOURce<hw>]:LIST:DWELl:LIST <Dwell>
Enters the dwell time values in the selected list in µs.
SOURce subsystem