EasyManuals Logo

Tektronix MSO56 User Manual

Tektronix MSO56
1026 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #121 background imageLoading...
Page #121 background image
Commands listed in alphabetical order
Examples
AFG:OUTP:LOA:
IMPED FIF
sets the AFG output load impedance to 50 Ohms.
AFG:OUTPUT:LOAD:IMPEDANCE? might return
:AFG:OUTPUT:
LOAD:IMPEDANCE HIGHZ
indicating the load impedance is set to
the high impedance state.
AFG:OUTPut
:STATE
Sets (or queries) the AFG o utput state.
Conditions
Requires option A FG.
Group
AFG
Syntax
AFG:OUTPut:STATE {0|1|OFF|ON}
AFG:OUTPut:STATE?
Arguments
1 or ON turns on the AFG output state.
0 or OFF turns it off.
Examples
AFG:OUTPUT:STATE ON turns on the AFG output state.
AFG:OUTPUT:STATE? might return :AFG:OUTPUT:STATE 1 indicating the AFG
output is on.
AFG:PERIod
Sets (or queries) the period of the AFG waveform, in seconds.
Conditions
Requires opti
on AFG.
Group
AFG
Syntax
AFG:PERIod <NR3>
AFG:PERIod?
Arguments
NR3 is the oating point number that represents the AFG period value, in seconds.
MSO54, MSO56, MSO58, MSO58LP Programmer 2-105