Commands listed in alphabetical order
TEKSecure (No
Query Form)
This command initializes, for the current user, both waveform and setup
memories, overwriting any previously stored data.
Equivalent to invoking Teksecure from the Utility menu. This is a time-consuming
operation (3 to 5 m inutes) and the instrument is inoperable until the TekSecure
operation i
s complete.
Group
Miscellaneous
Syntax
TEKSecure
Examples
TEKSECURE initializes both waveform and setup memories.
TIMe? (Query Only)
This command qu
eries the time that the instrument displays.
Group
Miscellaneous
Syntax
TIMe?
Related Commands
DATE?
Returns
<QString> isatimeintheform“hh:mm:ss”wherehhreferstoatwo-digithour
number, mm refers to a two-digit minute number from 01 to 60, and ss refers to a
two-digit second number from 01 to 60.
Examples
TIME?? might return :TIME "14:05:17", indicating the current time is set
to 2:05 p.m. and 17 seconds.
TIMe:ZONe
This command sets the time zone to the one specied.
Group
Miscellaneous
MSO54, MSO56, MSO58, MSO58LP Programmer 2-755

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix MSO56 and is the answer not in the manual?

Tektronix MSO56 Specifications

General IconGeneral
BrandTektronix
ModelMSO56
CategoryTest Equipment
LanguageEnglish

Related product manuals