Commands listed in alphabetical order
Group
Calibration
Syntax
CALibrate:INTERNal:STARt
Related Commands
CALibrate:INTERNal:STATus?
Examples
CALIBRATE:
INTERNAL:START
starts the signal path calibration.
CALibrate:INTERNal:STATus? (Query Only)
This query-only command returns the current status o f the signal path calibration.
NOTE. When running SPC through the remote interface, calibration status cannot
be obtained until after the SPC completes. SPC takes approximately
5 minutes per
channel which means a total of 40 minutes on an 8-channel model. Any remote
command that performs an action on the oscilloscope is also disabled until the
SPC is complete.
Group
Calibration
Syntax
CALibrate:INTERNal:STATus?
Related Commands
*CAL?
Returns
This query will return one of the following:
INIT indicates the instrument has not had signal p ath calibration run. The
instrument may need to be readjusted at the Tektronix service center.
PASS indicates that the signal path calibration co
mpleted successfully.
FAIL indicates that the signal path calibration did not complete successfully.
Examples
CALIBRATE:INTERNAL:STATUS? might return PASS , indicating that the current
status of the signal path calibration is that the signal path calibration completed
successfully.
CALibrate:PWRUpstatus? (Query Only)
This query-only command returns the current status of the power-up calibration.
MSO54, MSO56, MSO58, MSO58LP Programmer 2-187