EasyManua.ls Logo

Tektronix MSO56 - DISplay:WAVEView<x>:INTENSITy:GRATicule; DISplay:WAVEView<x>:INTENSITy:WAVEform

Tektronix MSO56
1026 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Commands listed in alphabetical order
Group
Display Contro
l
Syntax
DISplay:WAVEView<x>:GRAti cule {GRId|TIMe|FULl|NONe}
DISplay:WAVEView<x>:GRAti cule?
Arguments
GRId specifies a frame and grid only.
TIMe specifies a time graticule only.
FULl specifi es a frame, a grid and cross hairs.
NONe specified no graticule.
Examples
DISplay:WAVEView1:GRAticu le TIME specifies a time graticule.
DISplay:WAVEView1:GRAticu le? might return
:DISPLAY:WAVEVIEW1:GRATIC ULE GRID indicating the graticule is a g rid.
DISplay:WAVEView<x>:INTENSITy:GRATicule
This command sets or queries the graticule saturation level.
NOTE.
WAVEView<x>
is the specified waveview and must be
WAVEView1
.
Group
Display Control
Syntax
DISplay:WAVEView<x>:INTEN SITy:GRATicule <NR2>
DISplay:WAVEView<x>:INTEN SITy:GRATicule?
Arguments
<NR2> is the gra
ticule saturation level.
Examples
DISplay:WAVEView1:INTENSI Ty:GRATicule 75 sets the s aturation level
to 75%.
DISplay:WAVEView1:INTENSI Ty:GRATicule? might return
:DISPLAY:WAVEVIEW1:INTENS ITY:GRATICULE 66.0000 indicating the
saturation level is at 66%.
DISplay:WAVEView<x>:INTENSITy:WAVEform
This command sets or queries the waveform s
aturation level.
2-302 MSO54, MSO56, MSO58, MSO58LP Programmer