EasyManuals Logo

Tektronix MSO56 User Manual

Tektronix MSO56
1026 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #195 background imageLoading...
Page #195 background image
Commands listed in alphabetical order
Group
Bus
Syntax
BUS:B<x>:USB:DATAPLUSTHRES Hold <NR3>
BUS:B<x>:USB:DATAPLUSTHRES Hold?
Arguments
<NR3> is the P lus threshold.
Examples
BUS:B1:USB:DATAPLUSTHRESHo ld 50.0e-3 sets the threshold to 50.0 mV.
BUS:B1:USB:DATAPLUSTHRESHo ld? might return
:BUS:B1:USB:DATAPLUSTHRESH OLD 0.0E+0 indicating the threshold is
set to 0.0 V.
BUS:B<x>:USB:LOWTHRESHold
This command sets or queries the USB data source threshold for the specied bus
when the signal type is differential. The bus is specied by x.
Conditions
Requires option SR-USB2.
G
roup
B
us
Syntax
BUS:B<x>:USB:LOWTHRESHold <NR3>
BUS:B<x>:USB:LOWTHRESHold?
Arguments
<NR3> is the L ow threshold.
Examples
BUS:B1:USB:LOWTHRESHold 50.0e-3 sets the threshold to 50.0 mV.
BUS:B1:USB:LOWTHRESHold? might return :BUS:B1:USB :LOWTHRESHOLD
0.0E+0
indicating the threshold is set to 0.0 V.
BUS:B<x>:USB:SIGNALTYpe
This command sets or queries the USB signal type for the specied bus. The
bus is specied by x.
Conditions
Requires option SR-USB2.
MSO54, MSO56, MSO58, MSO58LP Programmer 2-179