Commands listed in alphabetical order
Arguments
FALL specifies t
he falling edge of the waveform.
RISE specifi es the rising edge of the waveform.
BOTH specifies both a rising and falling edge of the waveform.
SAMEas specifies that both edges of the waveform are the same.
OPPositeas specifies that the edges of the waveform a re not the same.
Examples
MEASUREMENT:MEAS3:DELAY:E DGE2 RISE specifies that the rising edge of the
waveform be used for measurement 3.
MEASUREMENT:MEAS1:DELAY:E DGE2? might return
:MEASUREMENT:MEAS1:DELAY: EDGE2 FALL, indicating that the falling edge
of the waveform is being us ed for measurement 1.
MEASUrement:MEAS<x>:DISPlaystat:ENABle
This command turns on and off display of statistics in measurement badges in the
user interface. This command affects only the display of statistics, basic-statistics
are computed regardless of the state of this command. Measurements are specified
by x.
Group
Measurement
Syntax
MEASUrement:MEAS<x>:DISPl aystat:ENABle {OFF|ON|<NR1>}
MEASUrement:MEAS<x>:DISPl aystat:ENABle
?
Arguments
OFF turns off the display of statistics in measurement badges.
ON turns on the display of statistics in measurement badges.
<NR1> = 0 turns o ff the display of statis
tics in the measurement badge, a ny other
value turns on the display of statistics.
Examples
MEASUrement:MEAS<x>:DISPl aystat:ENABle 0 turns off the display of
statistics in the measurement badge.
MEASUrement:MEAS<x>:DISPl aystat:ENABle? might return
:MEASUREMENT:MEAS1:DISPLA YSTAT:ENABLE 1 indicating that the display of
statistics in the measurement badge is on.
2-442 MSO54, MSO56, MSO58, MSO58LP Programmer