Commands listed in alphabetical order
DISplay:MATH
FFTView<x>:ZOOM:YAXIS:TO
This command sets or queries the top value of the zoom y axis in the specied
plot view.
Group
Zoom
Syntax
DISplay:MATHFFTView<x>:ZOO M:YAXIS:TO <NR3>
DISplay:MATHFFTView<x>:ZOO M:YAXIS:TO?
Arguments
<NR3> is the top value of the zoom y a xis in the specied plot view.
Examples
DISplay:MATHFFTView1:ZOOM: YAXIS:TO 21.0e0 sets the top value of the
zoom y axis in the specied plot view to 21 dBm.
DISplay:MATHFFTView1:ZOOM: YAXIS:TO? might return
:DISPLAY:MATHFFTVIEW1:ZOOM :YAXIS:TO 21.174171410663 3
indicating the top value of the zoom y axis in the specied plot view is 21.17 dBm.
DISplay:PERSistence
This command sets or queries the display persistence for analog waveforms.
Persistence is valid for wave views only.
Group
Display Control
Syntax
DISplay:PERSistence
{OFF|AUTO|INFPersist|INFIn ite|VARpersist|CLEAR}
DISplay:PERSistence?
Related Commands
DISplay:VARpersist
Arguments
OFF disables the persistence aspect of the display.
AUTO automatically set the persistence.
INFPersist se ts a display mode where any pixels, once touched by samples,
remain set until cleared by a mode change.
INFInite sets a display mode where any pixels, once touched by samples,
remain set until cleared by a mode change.
VARPersist sets a display mode where set pixels are gradually dimmed.
MSO54, MSO56, MSO58, MSO58LP Programmer 2-247

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix MSO56 and is the answer not in the manual?

Tektronix MSO56 Specifications

General IconGeneral
BrandTektronix
ModelMSO56
CategoryTest Equipment
LanguageEnglish

Related product manuals