Commands listed in alphabetical order
Related Commands
POWer:POWer<x
>:HARMONICS:STANDard
Argument
s
RATed: se l ect to use the standard input current values in the measurement.
MEASured: select to use the measured the input current values in the measurement.
Examples
POWer:POWer1:HARMONICS:CM EThod RATed sets the current method for
harmonics measurement as rated.
POWer:POWer<x>:HARMONICS:FUNDCURRent
This command sets o
r queries the fundamental current value for the harmonics
measurement in the specied power m easurement badge. The power measurement
badge is specied by x.
Conditions
Requires option 5-PWR, SUP5-PWR, or 5-PS2.
Group
Power
Syntax
POWer:POWer<x>:HARMONICS: FUNDCURRent <NR1>
POWer:POWer<x>:HARMONICS: FUNDCURRent?
Related Commands
POWer:POWer<x>:HARMONICS:CLASs
Arguments
<NR1> ranges from 0 to 16.
Examples
POWer:POWer<x>:HARMONICS: FUNDCURRent 1.5 sets the fundamental
current for the harmonics measurement as 1.5.
POWer:POWer<x>:HARMONICS:HORDer
This command sets or queries the order value for the harmonics measurement
in the specied power measurement badge. The power measurement badge
is specied by x.
Conditions
Requires option 5-PWR, SUP5-PWR, or 5-PS2.
Group
Power
2-538 MSO54, MSO56, MSO58, MSO58LP Programmer