Commands listed in alphabetical order
MEASUrement:
MEAS<x>:EDGE<x>
This command sets or queries the type of the specified edge, rise or fall, for the
measurement. The measurement number is specified by x.
Group
Measurement
Syntax
MEASUrement:MEAS<x>:EDGE<x > {RISE|FALL|BOTH}
Arguments
RISE specifies the rising edge.
FALL specifies the falling edge.
BOTH specifies either the rising or falling edge.
Examples
MEASUREMENT:MEAS2:EDGE1? might return :MEASUREMEN T:MEAS2:EDGE1
RISE
indicating that edge 1 of measurement 2 is the rising edge.
MEASUrement:MEAS<x>:FILTers:BLANKingtime
This command sets or queries the filter blanking time for the m easurement.
Measurements are specified by x.
Group
Measurement
Syntax
MEASUrement:MEAS<x>:FILTer s:BLANKingtime <NR3>
MEASUrement:MEAS<x>:FILTer s:BLANKingtime?
Arguments
<NR3> is the current filter blanking time.
Examples
MEASUrement:MEAS1:FILTers: BLANKingtime 3.5 sets the blanking time to
3.5.
MEASUrement:MEAS1:FILTers: BLANKingtime? might return
:MEASUREMENT:MEAS1:FILTERS :BLANKINGTIME 4.0000 indicating the filter
blanking time is 4.0.
MEASUrement:MEAS<x>:FILTers:GLOBal
This command sets or queries the global flag for filter settings for the
measurement. Measurements are specified by x.