Commands listed in alphabetical order
Syntax
POWer:POWer<x
>:SWITCHINGLOSS:ISOURce
{CH<x> | MATH<x> | REF<x>}
POWer:POWer<x>:SWITCHINGLO SS:ISOURce?
Arguments
CH<x> = A channel specifier; <x> is 1 through 8 and is limited by the number
of FlexChannels in your instrument.
MATH<x> = A math waveform specifier; <x> is ≥1.
REF<x> = A reference waveform specifier; <x> is ≥1.
Examples
POWer:POWer1:SWITCHINGLOSS :ISOURce CH2 sets the current source for the
switching loss measurement in the specified power measurement badge as CH2.
POWer:POWer2:SWITCHINGLOSS :ISOURce? might return CH1 indicating the
current source for switching loss measurement of Power 2 power measurement
badge.
POWer:POWer<x>:SWITCHINGLOSS:LEVELUNIts
This command sets or queries the level units for switching loss measurement
in the specified power measurement badge. The power measurement badge
is specified by x.
Conditions
Requires option 5-PWR, SUP5-PWR, or 5-PS2.
Group
Power
Syntax
POWer:POWer<x>:SWITCHINGLO SS:LEVELUNIts {PERCent | ABSolute}
POWer:POWer<x>:SWITCHINGLO SS:LEVELUNIts?
Arguments
PERCent to set the High, Mid, and Low reference levels in percentage.
ABSolute to set the High, Mid, and Low reference levels to s pecific signal levels.
Examples
POWer:POWer1:SWITCHINGLOSS :LEVELUNIts ABSolute sets the level units
as Absolute for switching loss measurement of the power measurement badge
Power 1.
MSO54, MSO56, MSO58, MSO58LP Programmer 2-589