EasyManuals Logo

Tektronix MSO56 User Manual

Tektronix MSO56
1026 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #392 background imageLoading...
Page #392 background image
Commands listed in alphabetical order
MATH:MATH<x>
:SPECTral:TYPE
This command sets or queries the FFT type selected for spectral analysis. The
math waveform is specied by x.
Group
Math
Syntax
MATH:MATH<x>:SPECTral:TYP E {MAGNitude|PHASe|REAL|IMAG inary}
MATH:MATH<x>:SPECTral:TYP E?
Arguments
MAGNitude species the magnitude spectral function.
PHASe species the phase spectral function.
REAL speci es the real spectral function.
IMAGinary species the imaginary spectral function.
Examples
MATH:MATH1:SPECTral:TYPE REAL species the real spectral function.
MATH:MATH1:SPECTral:TYPE? might return :MATH:MATH1:SPECTRAL:TYPE
MAGNITUDE
indicating the math is the magnitude spectral function.
MATH:MATH<x>:SPECTral:UNWRap
This command sets or queries whether phase unwrap of the spectral analyzer
output data is enabled. The Math waveform is specied by x.
Group
Math
Syntax
MATH:MATH<x>:SPECTral:UNW Rap {OFF|ON|0|1}
MATH:MATH<x>:SPECTral:UNW Rap?
Arguments
0 disables phase unwrap for the specied math waveform.
1 enables phase unwrap for the specied math waveform.
ON enables phase unwrap for the specied math waveform.
OFF disables phase unwrap for the spec ied math waveform.
Examples
MATH1:SPECTRAL:UNWRAP ON enables phase unwrap of the spectral a
nalyzer
output data.
2-376 MSO54, MSO56, MSO58, MSO58LP Programmer