Commands listed in alphabetical order
DVM:MEASUrem
ent:HIStory:MAXimum? (Query Only)
Returns the maximum readout value for the DVM f unction over the history
period. The history period is a constant period of 5 seconds.
Conditions
Requires DVM option.
Group
DVM
Syntax
DVM:MEASUrement:HIStory:M AXimum?
Examples
DVM:MEASU:HIS:MAX? might return :DVM:MEASUREMENT:HISTORY:MAXIMUM
431.9000E-3
, which represents the maximum readout value for the DVM
function over the history period in volts.
DVM:MEASUrement:HIStory:MINImum? (Query Only)
Returns the minimum readout value for the DVM over the history period. The
history period is a constant period of 5 seconds.
Conditions
Requires DVM option.
Group
DVM
Syntax
DVM:MEASUrement:HIStory:M INImum?
Examples
DVM:MEASU:HIS:MINI? might return
:DVM:MEASUREMENT:HISTORY: MINIMUM 430.9000E-3, which represents the
minimum readout value for the DVM function over the history period in volts.
DVM:MEASUrement:INFMAXimum? (Query Only)
Returns the maximum DVM readout value over the entire time that the DVM has
been on since the last change using the DVM:MODe or DVM:SOUrce commands
or DVM
RESET.
Conditions
Requires DVM option.
2-318 MSO54, MSO56, MSO58, MSO58LP Programmer

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix MSO56 and is the answer not in the manual?

Tektronix MSO56 Specifications

General IconGeneral
BrandTektronix
ModelMSO56
CategoryTest Equipment
LanguageEnglish

Related product manuals