Commands listed in alphabetical order
Syntax
WFMOutpre:XUN
it?
Related Commands
DATa:SOUrce
Examples
WFMOUTPRE:XUNIT? might return :WFMOUTPRE:XUNIT "HZ", indicating that
the horizontal units for the waveform are in Hertz.
WFMOutpre:XZEro? (Query Only)
This query-only command returns the sub-sample time between the trigger
sample (designated by PT_OFF) and the occurrence of the actual trigger for
the waveform specified by the DATa:SOUrce command. This value is in units
of WFMOutpre:XUNit.
An er
ror is reported if the DATa:SOUrce waveform does not exist.
NOTE. During steady state operation, when all control changes have settled and
triggers are arriving on a regular basis, this is the only part of the preamble that
changes on each acquisition.
Group
Waveform Transfer
Syntax
WFMOutpre:XZEro?
Related Commands
DATa:SOUrce, WFMOutpre:XUNit?
Examples
WFMOUTPRE:XZERO? might return :WFMOUTPRE:XZERO 5.6300E-9, indicating
that the trigger actually occurred 5.63 ns before the trigger sample.
WFMOutpre:YMUlt? (Query Only)
This query-only command returns the vertical scale factor per digitizing level
in units specified by WFMOutpre:YUNit for the waveform specified by the
DATa:SOUrce command. For those formats in which WFMOutpre:BYT_Nr is
important(allnon-floating point formats), WFMOutpre:YMUlt? must take the
location of the binary point implied by BYT_NR into consideration.
An error is reported if the DATa:SOUrce waveform does not exist.
Group
Waveform Transfer