Commands listed in alphabetical order
For Positive Du
ty Cycle measurement, the parameters is “PDUTY”.
For Negative Duty Cycle measurement, the parameters is “NDUTY”.
For Positive Pulse Width measurement, the parameters is “PPULSE”.
For Negative Pulse Width measurement, the p arameters is “NPULSE”.
For Cycle Amplitude measurement, the parameters is “AMPL”.
For Cycle Peak–Peak measurement, the parameters is “PKPK”.
For Cycle Top measurement, the parameters is “HIGH”.
For Cycle Base measurement, the parameters is “LOW”.
For Cycle Max measurement, the parameters is “MAX”.
For Cycle Min measurement , the parameters is “MIN”.
Examples
POWer:POWer1:RESUlts:CURRe ntacq:MINimum? "TruePwr" might
return 5.1307829019093E–9, indicating the minimum value of true power for all
acquisi
tions.
POWer:POWer<x>:RESUlts:ALLAcqs:PK2PK? (Query Only)
This command queries the peak-to-peak value of all acquisitions for the
measurement parameter in the specified power measurement badge. The pow
er
measurement badge is specified by x.
Conditions
Requires option 5-PWR, SUP5-PWR, or 5-PS2.
Group
Power
Syntax
POWer:POWer<x>:RESUlts:ALL Acqs:PK2PK? <QString>
Arguments
<QString> = parameters of the mea surements. The details of the parameters for
the Power measurements are listed below:
For Power Quality measurement, the parameters are "TRUEPWR" | "APPPWR"
| "REPWR" | "PWRFACTOR" | "PHASE" | "PWRFREQ" | "ICFACTOR" |
"VCFACTOR" | "IRMS" | "VRMS".
For Switching Loss measurement, the parameters are "TONENRG" | "TONLOSS"
| "TOFFENRG" | "TOFFLOSS" | "CONDENRG" | "CONDLOSS" | "TTLLOSS" |
"TTLENRG".
For dV by dt m easurement, the parameter is “DVBYDT”.
MSO54, MSO56, MSO58, MSO58LP Programmer 2-565