Commands listed in alphabetical order
Related Commands
TRIGger:{A|B}
:BUS:B<x>:ETHERnet:CONDition
TRIGger:{A|B}:BUS:B<x>:ETHERnet:TCPHeader:ACKnum:VALue
TRIGger:{A|B}:BUS:B<x>:ETHERnet:TCPHeader:SOUrceport:VALue
TRIGger:{A|B}:BUS:B<x>:ETHERnet:TCPHeader:DESTinationport:VALue
Arguments
<QString> is a quoted string of up to 3 2 characters where the allowable
characters are 0, 1, and X. The bits specified in the quoted string replace the least
significant bits, leaving any unspecified upper bits unchanged.
Examples
TRIGger:A:BUS:B1:ETHERnet :TCPHeader:SEQnum:VALue
"XXXXXXXXXXXXXXXXXXXX0001 00010001"
would set the sequence number
that is to be used in the trigger to hexadecimal
XXXXX111.
TRIGger:A:BUS:B1:ETHERnet :TCPHeader:SEQnum:VALue? might return
"XXXXXXXXXXXXXXXXXXXXXX00 10010100" .
TRIGger:{A|B}:BUS:B<x>:ETHERnet:TCPHeader:SOUrceport:VALue
When the Ethernet trigger condition is set to TCPHeader, this command specifies
the 16–bit source port address that is to be used in the trigger (along with the
destination port address, the sequence number and the acknowledgement number).
The default is all X’s (don’t care). The bus number is s
pecified by <x>.
Conditions
Requires the SR-ENET Triggering and Analysis application.
Group
Trigger
Syntax
TRIGger:{A|B}:BUS:B<x>:ET HERnet:TCPHeader:SOUrceport:VALue
<QString>
TRIGger:{A|B}:BUS:B<x>:ET HERnet:TCPHeader:SOUrceport:VALue?
Related Commands
TRIGger:{A|B}:BUS:B<x>:ETHERnet:CONDition
TRIGger:{A|B}:BUS:B<x>:ETHERnet:TCPHeader:ACKnum:VALue
TRIGger:{A|B}:BUS:B<x>:ETHERnet:TCPHeader:SEQnum:VALue
TRIGger:{A|B}:BUS:B<x>:ETHERnet:TCPHeader:DESTinationport:VALue
Arguments
<QString> is a quoted string of up to 1 6 characters where the allowable
characters are 0, 1, and X. The bits specified in the quoted string replace the least
significant bits, leaving any unspecified upper bits unchanged.
2-792 MSO54, MSO56, MSO58, MSO58LP Programmer