Commands listed in alphabetical order
Arguments
EXECUTE starts
execution of the diagnostics.
ABOrt disables diagnostics capabilities and returns the instrument to a n ormal
operating sta
te.
Examples
DIAG:STATE ABORT turns off diagnostics capabilities and returns the instrument
to a normal operating state.
DIAG:STATE? might return :DIA G:STATE ABORT, indicating that diagnostics
are disabled.
DIGGRP<x>:D<x>:THReshold
Sets or queries the threshold level in volts for the specied digital channel. If the
source channel doesn't exist, a hardware missing error event is set.
Group
Digital
Syntax
DIGGRP<x>:D<x>:THReshold <NR3>
DIGGRP<x>:D<x>:THReshold?
Arguments
<NR3> is the threshold level in volts for the specied digital channel.
Examples
DIGGRP1:D1:THReshold 30.0E-3 sets the threshold to 30 mV.
DIGGRP1:D1:THReshold? might return :DIGGRP1:D1:THRESHOLD 0.0E+0
indicating the threshold is 0.0 V.
DIGGRP<x>:THReshold
This command sets or queries the digital threshold of all bits on the specied
digital channel group. The digital channel group is specied by <x>.
Group
Digital
Syntax
DIGGRP<x>:THReshold <NR3>
DIGGRP<x>:THReshold?
Arguments
<NR3> is the digital threshold of all bits on DCH<x>.
2-226 MSO54, MSO56, MSO58, MSO58LP Programmer