EasyManuals Logo
Home>Tektronix>Test Equipment>MSO56

Tektronix MSO56 User Manual

Tektronix MSO56
1026 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #888 background imageLoading...
Page #888 background image
Commands listed in alphabetical order
Examples
TRIGGER:A:UPP
ERTHRESHOLD:CH1 1.3
This command sets the A trigger level
for Channel 1 to 1.3 V level.
TRIGGER:A:UPPERTHRESHOLD: CH2? might return
:TRIGGER:A:UPPERTHRESHOLD :CH2 1.3000E+00, indicating that the A
trigger level for Channel 2 is set to 1.3 V.
TRIGger:{A|B}:WINdow:CROSSIng
This command sets or queries the window trigger threshold crossing of the
selected trigger source. The threshold crossing selection is only effective when
:TRIGger:{A|B}:WINdow:WHEn is INSIDEGreater or OUTSIDEGreater.
Group
Trigger
Syntax
TRIGger:{A|B}:WINdow:CROS SIng {UPPer|LOWer|EITher|NON e}
Arguments
UPPer if :TRIG ger:{A|B}:WINdo w:WHEn is INSIDEGreater, the instrument
triggers when the signal remains between the up
per and lower thresholds for longer
than the time limit (
:TRIGger:{A|B}:WINdow:WIDTH) and then exits through
the upper threshold. If
:TRIGger:{A|B}:WINdow:WHEn is OUTSIDEGreater,
the instrument triggers when the signal remains above the upper threshold for
longer than the time limit (
:TRIGger:{A|B}:WINdow:WIDT H) and then crosses
downward through the upper threshold.
LOWer if :TRIG ger:{A|B}:WINdo w:WHEn is INSIDEGreater, the instrument
triggers when the signal remains between the upper and lower thresholds for longer
than the time limit (
:TRIGger:{A|B}:WINdow:WIDTH) and then exits through
the lower threshold. If
:TRIGger:{A|B}:WINdow:WHEn is OUTSIDEGreater,
the instrument triggers when the signal remains below the lower threshold for
longer than the time limit (
:TRIGger:{A|B}:WINdow:WIDT H) and then crosses
upwards through the lower threshold.
EITher if :TRIGger:{A|B}:WINdow:WHEn is INSIDEGreater, the instrument
triggers when the signal remains between the upper and lower thresholds for
longer than the time limit
(
:TRIGger:{A|B}:WINdow:WIDTH) and then exits
through either the upper or lower threshold. If
:TRIGger:{A|B}:WINdow:WHEn
is OUTSIDEGreater, the instrument triggers when the signal remains either
above the upper threshold or below the lower threshold for longer than the time
limit (
:TRIGger:{A|B}:WINdow:WIDTH) and then crosses a threshold.
NONe if :TRIGger:{A|B}:WINdow:WHE n is INSIDEGr eater, the instrument
triggers when the signal remains between the upper and lower thresholds for
longer than the time limit (
:TRIGger:{A|B}:WINdow:WIDT H) without crossing
through either the upper or lower threshold. If
:TRIGger:{A|B}:WINdow:WHEn
is OUTSIDEGreater, the instrument triggers when the signal remains
2-872 MSO54, MSO56, MSO58, MSO58LP Programmer

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix MSO56 and is the answer not in the manual?

Tektronix MSO56 Specifications

General IconGeneral
BrandTektronix
ModelMSO56
CategoryTest Equipment
LanguageEnglish

Related product manuals