Commands listed in alphabetical order
Group
Measurement
Syntax
MEASUrement:MEAS<x>:CLOCK Recovery:MODel {TYPE1|TYPE2}
MEASUrement:MEAS<x>:CLOCK Recovery:MODel?
Arguments
Argumenst are the clock recovery model type.
Examples
MEASUrement:MEAS1:CLOCKRe covery:MODel TYPE1 set the clock recove ry
model lto type 1.
MEASUrement:MEAS1:CLOCKRe covery:MODel? might return
:MEASUREMENT:MEAS1:CLOCKR ECOVERY:MODEL TYPE1 indicating the clock
recovery model is type 1.
MEASUrement:MEAS<x>:CLOCKRecovery:NOMINALOFFset
This command sets or queries the offset value used when explicit clock recovery
is used for the measurement. Measurements are specied by x.
Group
Measurement
Syntax
MEASUrement:MEAS<x>:CLOCK Recovery:NOMINALOFFset <N R3>
MEASUrement:MEAS<x>:CLOCK Recovery:NOMINALOFFset?
Arguments
<NR3> is the clock offset.
Examples
MEASUrement:MEAS1:CLOCKRecovery:NOMINALOFFset 1.0e-9 sets the
clock offset to 1 ns.
MEASUrement:MEAS1:CLOCKRecovery:NOMINALOFFset? might return
:MEASUREMENT:MEAS1:CLOCKRECOVERY:NOMINALOFFSET 0.0E+0
indicating the offset is 0.0 ns.
MEASUrement:MEAS<x>:CLOCKRecovery:NOMINALOFFset:SELECTIONtype
This command sets or queries the offset type used when explicit clock recovery is
used for the measurement. The measurement number is specied by x.
Group
Measurement
2-438 MSO54, MSO56, MSO58, MSO58LP Programmer