EasyManuals Logo

Tektronix MSO56 User Manual

Tektronix MSO56
1026 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #372 background imageLoading...
Page #372 background image
Commands listed in alphabetical order
Examples
HORizontal:SA
MPLERate:ANALYZemode:MINimum :VALue AUTOMATIC
allows the oscilloscope to set the minimum value.
HORizontal:S
AMPLERate:ANALYZemode:MINi mum:VALue?
might return
:HORIZONTAL:SAMPLERATE:AN ALYZEMODE:MINIMUM:VALUE A UTOMATIC
indicating the oscilloscope automatically sets the minimum value.
HORizontal
:SCAle
This command sets or queries the horizontal scale.
Group
Horizontal
Syntax
HORizontal:SCAle <NR3>
HORizontal:SCAle?
Arguments
<NR3> is the horizontal scale in time per division..
Returns
The current horizontal scale is returned.
Examples
HORIZONTAL:SCALE 20e-9 sets the horizontal scale to 20 ns/division.
HORIZONTAL:SCALE? might return :HORIAONTAL:SCALE 20e-9 indicating the
horizontal scale is set to 20 ns/division.
ID? (Query Only)
This query-only command returns identifying information about the instrument
and related rmware similar to that returned by the *IDN? IEEE488.2 common
query but does not include the instrument serial number.
Group
Miscellaneous
Syntax
ID?
Related Commands
*IDN?
2-356 MSO54, MSO56, MSO58, MSO58LP Programmer

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix MSO56 and is the answer not in the manual?

Tektronix MSO56 Specifications

General IconGeneral
BrandTektronix
ModelMSO56
CategoryTest Equipment
LanguageEnglish

Related product manuals