EasyManuals Logo

Tektronix MSO56 User Manual

Tektronix MSO56
1026 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #518 background imageLoading...
Page #518 background image
Commands listed in alphabetical order
MEASUrement:
REFLevels:MODE
This command sets or queries how often reference levels are calculated.
Group
Measurement
Syntax
MEASUrement:REFLevels:MOD E {LATCh|CONTinuous}
MEASUrement:REFLevels:MOD E?
Arguments
LATCh calculate s reference levels only on the rst acquisition after a statistics
reset.
CONTinuous calculates reference levels on every acquisition.
Examples
MEASUREMENT:REFLEVELS:MOD E CONTINUOUS calculates reference levels on
every acquisition.
MEASUREMENT:REFLEVELS:MOD E? might return
:MEASUREMENT:REFLEVELS:MO DE LATCH indicating reference levels are
calculated only on the rst acquisition after a statistics reset.
MEASUrement:REFLevels:PERCent:FALLHigh
This command sets or queries the percentage (where 100% is equal to TOP and
0% is equal to BASE) used to calculate the high reference level of the falling edge
when the measurement's ref level method is set to percent.
Group
Measurement
Syntax
MEASUrement:REFLevels:PER Cent:FALLHigh <NR3>
MEASUrement:REFLevels:PER Cent:FALLHigh?
Arguments
<NR3> is the percentage used to calculate the high reference level of the falling
edge
Examples
MEASUrement:REFLevels:PER Cent:FALLHigh 90 sets the h igh reference
level to 90%.
MEASUrement:REFLevels:PER Cent:FALLHigh? might return
:MEASUREMENT:REFLEVELS:PE RCENT:FALLHIGH 80.0000 indicating the
high reference level is 80%.
2-502 MSO54, MSO56, MSO58, MSO58LP Programmer

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix MSO56 and is the answer not in the manual?

Tektronix MSO56 Specifications

General IconGeneral
BrandTektronix
ModelMSO56
CategoryTest Equipment
LanguageEnglish

Related product manuals