Commands listed in alphabetical order
Group
Measurement
Syntax
MEASUrement:MEAS<x>:MEASRa nge:GLOBal {OFF|ON|0|1}
MEASUrement:MEAS<x>:MEASRa nge:GLOBal?
Arguments
OFF species that range settings can be set independently for each individual
measurement.
ON applies global measurement range settings to all the measurements' range
settings.
0 species that range settings can be set independently for each individual
measurement.
1 applies global measurement range settings to all the measurements' range
settings.
Examples
MEASUrement:MEAS1:MEASRang e:GLOBal 1 applies global measurement
range settings to all the measurements' range settings.
MEASUrement:MEAS1:MEASRang e:GLOBal? might return
:MEASUREMENT:MEAS1:MEASRAN GE:GLOBAL 0 indicating that range settings
can b e s et independently for each individual measurement.
MEASUrement:MEAS<x>:MEASRange:MAX
This command sets or queries the range maximum value for the measurement.
Measurements are specied by x.
Group
Measurement
Syntax
MEASUrement:MEAS<x>:MEASRa nge:MAX <NR3>
MEASUrement:MEAS<x>:MEASRa nge:MAX?
Arguments
<NR3> is the maximum measurement range limit value.
Examples
MEASUrement:MEAS1:MEASRang e:MAX 2.50 sets the maximum range to 2.5 V.
MEASUrement:MEAS1:MEASRang e:MAX? might return
:MEASUREMENT:MEAS1:MEASRAN GE:MAX 1.0000 indicating the
maximum range is 1.0 V.
MSO54, MSO56, MSO58, MSO58LP Programmer 2-463